9 free EMT practice questions: Medication & Supply Inventory
These are real questions from the same bank the app draws from. Each one is written to the NREMT EMT content specifications and kept inside the EMT scope of practice. Pick an answer and you get the full rationale, including why the other three options are wrong.
Work through all 9, then move on to the next topic. When you want the full picture, the free EMT diagnostic covers every topic in one sitting. No account needed for any of it.
Question 1 of 9
During a supply check on a unit assigned to cover a summer day camp, an EMT finds that the only epinephrine auto-injectors carried are the 0.3 mg strength. What is the concern with this finding?
Show the answer and rationale
Correct answer · A small child could not be given the correct dose of epinephrine
A small child could not get the right dose of epinephrine. This unit is covering a summer day camp, and most of the patients at a day camp are small children. The unit only carries 0.3 mg auto-injectors. An auto-injector gives one fixed dose, and you cannot split the dose. The dose for a small child, under about 66 pounds, is 0.15 mg. Giving the 0.3 mg adult dose to a small child puts twice the right dose into the child. Too much epinephrine pushes the heart rate and blood pressure up too far. Stock 0.15 mg pediatric auto-injectors before this unit covers the camp.
Why the others are wrong
Epinephrine auto-injectors expire sooner than other medications: Checking expiration dates is part of every supply check, and epinephrine auto-injectors do have a shelf life. Nothing here says these devices are expired or close to it. The problem is the strength on the unit, not the date.
The 0.3 mg strength may be given only by an advanced provider: Some medications on a unit are limited to advanced providers. Epinephrine auto-injectors are not one of them. An EMT can give either the 0.3 mg or the 0.15 mg auto-injector to a patient in anaphylaxis.
Two auto-injectors would be needed for every adult patient: The 0.3 mg dose can look small, and stacking two devices sounds like a way to reach an adult dose. The 0.3 mg auto-injector is already the full adult dose. One device gives an adult the whole dose. Some adults do need a repeat dose later, and each repeat is another full 0.3 mg dose, not half of one.
Question 2 of 9
During a daily supply check, an EMT finds that the ambulance carries nasal cannulas and BVMs but no NRBs. Which capability does the unit lack until the masks are restocked?
Show the answer and rationale
Correct answer · Delivery of high-concentration oxygen to an adequately breathing patient
Without nonrebreather masks, the unit cannot give high-concentration oxygen to a patient who is breathing adequately. A nonrebreather runs at 10 to 15 L/min and delivers roughly 60 to 90% oxygen. A nasal cannula runs at 1 to 6 L/min and delivers roughly 24 to 44%, which makes the cannula a low-flow device. A BVM with a reservoir gets close to 100%, but the BVM is for ventilating a patient whose breathing is inadequate. With the nonrebreathers gone, you can still give low-flow oxygen, and you can still ventilate. The only thing missing is high-concentration oxygen for a patient breathing on their own.
Why the others are wrong
Delivery of low-flow oxygen to an adequately breathing patient: Low-flow oxygen for a patient who is breathing adequately comes from the nasal cannula. The unit still has nasal cannulas.
Measurement of the oxygen saturation of an adequately breathing patient: Oxygen saturation is measured with a pulse oximeter. Which masks are stocked has nothing to do with the pulse oximeter.
Ventilation of a patient whose breathing is inadequate: A patient whose breathing is inadequate gets ventilated with a BVM, and the BVMs are still on the unit.
Question 3 of 9
During a shift check at a station that covers an elementary school, an EMT finds that the automated external defibrillator case contains adult pads only. What is the correct interpretation of this finding?
Show the answer and rationale
Correct answer · Pediatric pads should be stocked, but adult pads may be used if needed
Pediatric pads need to be stocked, and adult pads can still be used on a child if needed. This station covers an elementary school, and the AED case has adult pads only. Pediatric pads have a dose attenuator that lowers the energy of the shock. The lower dose keeps a child's small heart from getting an adult-sized shock. Pediatric pads are the standard for children under 8 years old or under 55 pounds. Missing pediatric pads are a stocking problem to fix. Report the missing pads so the unit gets restocked. If a child goes into cardiac arrest before the pediatric pads arrive, defibrillate with the adult pads.
Why the others are wrong
Pediatric pads are unnecessary, because adult pads fit every patient: Adult pads can be used on a child when nothing else is available. Pediatric pads exist to lower the energy delivered to a child under 8 years old or under 55 pounds. A station that covers an elementary school needs to be ready for those children.
The defibrillator may not be used on a child until pediatric pads arrive: Pediatric pads are preferred for a child, but they are not required to shock. Guidelines say to use adult pads instead of holding back the shock. Waiting for pediatric pads delays lifesaving care for a child in cardiac arrest.
Pediatric pads are needed only for patients under 1 year of age: Pediatric pads are not only for infants. The cutoff is 8 years old or 55 pounds, well past the first year of life.
Question 4 of 9
During the morning equipment check, the EMT discovers that the epinephrine auto-injectors expire in 4 weeks. What is the appropriate action?
Show the answer and rationale
Correct answer · Document the expiration date and initiate replacement before they expire
Document the expiration date and start the replacement now. The auto-injectors expire in 4 weeks. An expiration date is the manufacturer's guarantee that the drug is at full strength. Your goal is to never have that date arrive while the auto-injectors are still on the truck. Writing down the date means the date gets tracked instead of remembered. Starting the replacement now gives the new stock time to arrive before the old stock expires. The truck then has in-date epinephrine for an anaphylaxis call in week five.
Why the others are wrong
Continue using them until they expire; replacement is not needed until after expiration: Using the auto-injectors right up to the expiration date is fine when the replacements are already on the way. The drug is at full strength until that date, and pulling it early wastes good stock. The problem is the second half of option A, waiting until after expiration to replace them. Waiting leaves a gap where the truck carries expired auto-injectors or none at all.
Inspect them for visible changes and use them if they appear normal: Looking at the auto-injectors is a real check you have to do. You look through the window for discoloration or particles. Cloudy or brown epinephrine gets pulled no matter what the date is. Looking at the drug cannot tell you its strength, because epinephrine breaks down chemically before it changes how it looks. Inspection and the expiration date are two separate checks, and one does not replace the other.
Discard them immediately and report the supply shortage to administration: Throwing out a medication right away is right when the medication is already expired, damaged, discolored, or recalled. These auto-injectors are good for 4 more weeks. Throwing them out now leaves the truck with no epinephrine and creates the shortage you would be reporting.
Question 5 of 9
During the morning equipment check, an EMT notices that one of the aspirin bottles is open and has a faint smell of moisture. The tablets look slightly discolored compared to a new bottle. What is the most appropriate action?
Show the answer and rationale
Correct answer · Remove the bottle from service immediately, document the issue, and ensure it is replaced with a new supply
This aspirin has been damaged by moisture and has to come out of service. The bottle is open, there is a damp smell, and the tablets look discolored next to a new bottle. Aspirin is acetylsalicylic acid. Water breaks aspirin down into salicylic acid and acetic acid. That's why damp aspirin picks up a faint vinegar smell. Damaged aspirin no longer has the strength printed on the label, and you count on that strength for a chest pain patient. Pull the bottle out of service now. Document what you found and when. Get the bottle replaced so the next crew does not reach for damaged aspirin on an ACS call.
Why the others are wrong
Recap the bottle, add a desiccant packet, and return it to service with a note to monitor: A desiccant packet and a tight cap prevent moisture damage. That's why the packet and the cotton are in the bottle in the first place. Once the tablets have absorbed water and started to break down, drying the air around them cannot undo the damage or bring back the strength. Returning the bottle to service with a note also leaves a bad drug on the truck. The next person to use the bottle may never read the note.
Use the tablets for patients who refuse liquid medication; discard them if the patient declines: Matching the form of a medication to what a patient will take is real clinical thinking, like chewable versus swallowed. The form has nothing to do with whether the drug is safe to give. Damaged medication is off limits for every patient, no matter what the patient prefers. Discarding the tablets only when a patient declines still means giving damaged aspirin to any patient who accepts.
Assume the discoloration is from normal storage and use the bottle until the expiration date passes: The expiration date is the usual reason to pull stock, and checking dates is part of the same rig check that caught this bottle. The printed date assumes the drug was stored right: sealed, dry, and within its temperature range. This bottle was not stored that way. Moisture can ruin a drug well before the printed date.
Question 6 of 9
While restocking at the station, an EMT carries a full spare oxygen cylinder to the ambulance and finds that the fixed housing that holds the spare has broken loose from the wall and is awaiting repair. No other bracket or commercially manufactured securing device is available on the unit. The unit is due back in service. Which action is most appropriate?
Show the answer and rationale
Correct answer · Leave the spare at the station and report the broken housing
Leave the spare cylinder at the station and report the broken housing. The unit is supposed to carry a spare oxygen cylinder, so leaving the spare behind does leave a gap. The safety rule for securing tanks comes first. Portable oxygen tanks have to be held by fixed clasps or housings, no matter where on the unit they ride. A tank can be secured to the stretcher or bench only with a commercially made device built for that job. This unit does not have one. A loose cylinder becomes a projectile in a crash. The cylinder's valve and gauge can also be damaged if the cylinder falls. Write the broken housing on the shift checklist so the housing gets documented and fixed.
Why the others are wrong
Strap the cylinder to the stretcher frame with the patient straps: A strong strap seems like it should hold a cylinder. Securing a tank to the stretcher is only allowed with a commercially made device built for that purpose. The question says the unit has no such device. Patient straps are made to hold a patient, and the stretcher itself moves in and out of the ambulance.
Lay the cylinder on the floor behind the airway seat, out of the walkway: Laying the cylinder on the floor out of the walkway keeps people from tripping on it. Tripping is not the main danger. A tank that is not secured becomes a projectile in a crash, wherever the tank starts out in the compartment.
Stow the cylinder in the outside compartment with the extrication tools: Moving the cylinder to an outside compartment keeps the cylinder away from the crew. The rule to secure tanks with fixed clasps or housings applies wherever they are carried. Loose extrication tools in the same compartment can also damage the cylinder's valve and gauge.
Question 7 of 9
An EMT is checking the spare portable oxygen cylinders carried on the ambulance. One aluminum cylinder is labeled for medical oxygen, its plastic valve seal was intact until the EMT removed it, and the regulator gauge reads 2,000 psi once the regulator is attached. The month and year stamped into the collar show that the cylinder was last tested 6 years ago. Which action is most appropriate?
Show the answer and rationale
Correct answer · Remove it from service and report it for retesting
This cylinder is overdue for testing. The month and year stamped into the collar show the cylinder was last tested 6 years ago. Aluminum cylinders are usually tested every 5 years, and composite cylinders every 3 years. Either way, 6 years is past due. The test is what shows the cylinder can still safely hold its pressure. Once the cylinder is overdue, nobody knows whether the cylinder is still safe, no matter what the gauge reads. The label, the seal, and the gauge each check something else. The label tells you the cylinder holds medical oxygen. The intact seal tells you nobody opened the cylinder since it was filled. The gauge tells you how much oxygen is inside. Take the cylinder off the unit and send it back for testing. Write the overdue cylinder on the shift checklist so a replacement is on the ambulance before the unit goes in service.
Why the others are wrong
Place it in service, since the gauge reads full: The gauge reading of 2,000 psi tells you how much oxygen is inside. The collar stamp is what tells you when the cylinder was last tested. A cylinder can read 2,000 psi and still be overdue for the test that shows it can safely hold that pressure.
Crack the valve to confirm it holds pressure: Cracking the valve is a real step. Cracking the valve clears debris from the opening before the regulator goes on. Holding pressure for a few seconds on the station floor is not the test the collar stamp records.
Keep it as the backup and use the newer one: The backup is the cylinder you reach for with a second patient or a long transport. Moving the overdue cylinder to backup keeps it on the ambulance. You still end up depending on the overdue cylinder, just later.
Question 8 of 9
An EMT is restocking the ambulance after a 90-minute transfer between two hospitals during which the mounted oxygen system was used with a disposable humidifier. The humidifier bottle still holds most of its sterile water and shows no visible contamination. The next call may be another long transport. Which action is most appropriate?
Show the answer and rationale
Correct answer · Discard the humidifier and fit a new single-use bottle
Humidifier equipment is single use. You throw the humidifier away after one patient so the bottle cannot carry infection to the next patient. This bottle was used on a patient during the 90-minute transfer, so it is done. Most of the sterile water is still there, and the bottle shows no visible contamination. Neither one matters. The problem is the germs a used bottle can carry, not how full the bottle is or how clean it looks. Discard the bottle and fit a new single-use humidifier. A humidifier is considered for transports longer than 1 hour and usually is not needed on shorter ones. With a new bottle in place, the unit is ready if the next call is another long transport.
Why the others are wrong
Top off the sterile water and leave it in place: Topping off the water keeps the same used bottle in the oxygen line for the next patient. Humidifier equipment is used on one patient only. The bottle is what has to be replaced, not just the water.
Empty and air-dry the bottle before the next patient: Emptying and air-drying is how reusable equipment gets handled. A single-use humidifier does not go back into service after any cleaning step. This bottle has already been used on a patient, so it gets replaced.
Keep the bottle for the next transport over 1 hour: The 1-hour rule tells you whether to use a humidifier at all. The rule does not tell you how many patients one bottle can serve. A second long transport means a second patient, and that patient gets a new bottle.
Question 9 of 9
During a shift check, an EMT finds that the cervical collars stocked on the ambulance are all one size. What is the concern with this finding?
Show the answer and rationale
Correct answer · A single size cannot properly fit every patient who may need one
One cervical collar size cannot properly fit every patient. Keeping the ambulance stocked means having the range of sizes each piece of equipment needs to be used correctly. Patients have different neck sizes. With only one collar size, you cannot do spinal motion restriction correctly on the range of patients you may see. Restock the collars in the full range of sizes.
Why the others are wrong
There is no concern, since one size fits all adult patients: One size does not fit all adults. Adult necks vary in size, and a single collar size cannot properly fit every one of them.
Cervical collars are not required equipment on an ambulance: Cervical collars are standard spinal motion restriction equipment. They should be on the ambulance in the right sizes.
The concern only applies if a pediatric patient is anticipated: The problem is not only with children. Even among adults, one collar size cannot fit the full range of neck sizes you will see.
Find out which EMT topics are costing you points
Ten questions on one topic tell you about that topic. The free diagnostic covers every EMT topic and breaks your results down by topic, so you know what to drill next. No card, no signup to try it.
Take the free EMT diagnosticMore free EMT practice questions by topic
Scene Size-up and Safety
Primary Assessment
Secondary Assessment
Patient Treatment and Transport
- Cardiovascular Emergencies & Cardiac Arrest
- Respiratory Emergencies & Airway Management
- Shock (All Types) & Hemorrhage Control
- Neurological & EENT Emergencies
- Endocrine Disorders
- Abdominal, GI, GU & Renal Disorders
- Hematologic, Immunologic & Infectious Disorders
- Psychiatric & Behavioral Emergencies
- Toxicology & Substance Use Emergencies
- Head, Spinal & Nervous System Trauma
- Chest, Abdominal & Multi-System Trauma
- Orthopedic, Soft-Tissue & Burn Trauma
- Obstetric, Neonatal & Gynecologic Emergencies
- Pediatric, Geriatric & Special Populations
- Medication Administration
- Environmental Emergencies
Operations
- Equipment Readiness & Ambulance Check
- Medication & Supply Inventory
- Documentation (PCR)
- Responder Wellness & Stress Management